Fuel retention in plasma-facing components affects the operational efficiency and safety of fusion devices. However, detection of retained fuel by laser-induced breakdown spectroscopy (LIBS) remains challenging because the Hα and Dα Balmer lines commonly used for hydrogen-isotope identification are separated by only about 0.18 nm, making their discrimination highly sensitive to fusion-relevant ambient conditions. In this study, LIBS was employed to systematically investigate Hα/Dα discrimination on real EAST-exposed graphite tiles under various simulated ambient conditions. Depth-resolved measurements showed that the second laser pulse generally provided a more favorable Hα/Dα emission ratio and improved fitting quality, with the noise-normalized root mean square residual decreasing from 3.6 to 1.26 compared with the first pulse. Meanwhile, time-resolved plume analysis helped explain the role of ambient environments in governing spectral broadening and isotope discrimination. At 100 Pa, the initial plume-height expansion rates within 200 ns were 15.6, 14.2, and 8.0 km s−1 in He, air, and Ar, respectively, revealing pronounced gas-dependent differences in plasma expansion. Combined spectral and plume analyses showed that vacuum provided high spectral resolution but suffered from a relatively short plasma lifetime. Within the investigated ambient parameters, a helium atmosphere at a moderate pressure of 100 Pa offered the most favorable balance between plasma lifetime and spectral resolution. In contrast, higher pressures or denser gases tended to enhance Stark broadening, leading to more severe peak overlap and a narrower effective discrimination window. The effective discrimination window was evaluated using the resolution factor Rs, with Rs≈1.5 taken as a reference for baseline separation. These findings provide practical guidance for LIBS-based fuel retention diagnostics under EAST-relevant conditions.
プラズマ対向機器における燃料保持は、核融合装置の運転効率と安全性に影響を与える。しかし、レーザー誘起絶縁破壊分光法(LIBS)による保持燃料の検出は、水素同位体同定に一般的に用いられるHα線とDα線のバルマー線の間隔がわずか約0.18 nmしかなく、その識別が核融合関連の周囲環境条件に非常に敏感であるため、依然として困難である。本研究では、LIBSを用いて、実際のEAST曝露黒鉛タイル上のHα/Dα識別を、様々な模擬周囲環境条件下で系統的に調査した。深さ分解測定により、2回目のレーザーパルスは一般に、より好ましいHα/Dα発光比と改善されたフィッティング品質をもたらし、ノイズ正規化二乗平均平方根残差は1回目のパルスと比較して3.6から1.26に減少した。一方、時間分解プルーム解析は、スペクトル幅広がりと同位体識別を支配する周囲環境の役割の説明に役立った。100 Paでは、200 ns以内の初期プルーム高さ膨張速度は、He、空気、Ar中でそれぞれ15.6、14.2、8.0 km s−1であり、プラズマ膨張における顕著なガス依存性の差異が明らかになった。スペクトル解析とプルーム解析を組み合わせることで、真空は高いスペクトル分解能を提供するが、プラズマ寿命が比較的短いことが示された。調査した周囲環境パラメータの範囲内では、100 Paの中程度の圧力のヘリウム雰囲気が、プラズマ寿命とスペクトル分解能の間で最も好ましいバランスを提供した。対照的に、より高い圧力またはより高密度のガスは、シュタルク幅広がりを増強する傾向があり、より深刻なピーク重なりとより狭い実効識別ウィンドウをもたらした。実効識別ウィンドウは分解能因子Rsを用いて評価され、ベースライン分離の基準としてRs≈1.5が採用された。これらの知見は、EAST関連条件下でのLIBSベースの燃料保持診断に実用的な指針を提供する。