AbstractThe analytical gas chromatographic system (AN-GC) was the main tool used to analyse gas mixtures from various subsystems of the Active Gas Handling System (AGHS) and to characterise the functioning of the different processes during the Deuterium–Tritium Experiment (DTE1). Samples were transferred to AN-GC via long 14′′ lines. Calibration factors, accuracy and minimum detection limits for different gas species were determined for the thermal conductivity detectors, flame ionisation detectors and ionisation chambers. The performance of the AN-GC is discussed by means of various chromatograms and the experience gained with these detectors as well as their limits are mentioned. The AN-GC worked very reliably and about 320 analyses were performed during DTE1 and the subsequent cleanup phase.
分析用ガスクロマトグラフィーシステム(AN-GC)は、アクティブガス処理システム(AGHS)の様々なサブシステムからのガス混合物を分析し、重水素–トリチウム実験(DTE1)中の異なるプロセスの機能を特性評価するための主要なツールであった。試料は長さ14インチのラインを介してAN-GCに移送された。熱伝導度検出器、水素炎イオン化検出器、および電離箱について、異なるガス種に対する検量係数、精度、および検出下限が決定された。AN-GCの性能は、様々なクロマトグラムを用いて考察され、これらの検出器から得られた経験とその限界についても言及されている。AN-GCは非常に高い信頼性で動作し、DTE1中およびその後のクリーンアップ段階において約320回の分析が実施された。